Single channel 마이크로피펫 (Gilson) | |||||||
PIPETMAN? P Range of Models (Single channel) | |||||||
|
Product Name |
마이크로피펫 |
Maximum Permissible Errors |
| |||
|
Model |
Single channel |
Gilson |
ISO 8655 |
| ||
|
(Reference) |
Volume(㎕) |
systematic (μL) |
Random (μL) |
systematic (μL) |
Random (μL) |
|
|
F144801 (P2) |
Min 0.2 |
± 0.024 |
≤0. 012 |
± 0.08 |
≤0.04 |
|
|
0.5 |
± 0.025 |
≤0. 012 |
± 0.08 |
≤0.04 |
| |
|
Max 2 |
± 0.030 |
≤0. 014 |
± 0.08 |
≤0.04 |
| |
|
F144802 (P10) |
Min. 1 |
± 0.025 |
≤0. 012 |
± 0.12 |
≤0.08 |
|
|
5 |
± 0.075 |
≤0. 03 |
± 0.12 |
≤0.08 |
| |
|
Max 10 |
± 0.100 |
≤0. 04 |
± 0.12 |
≤0.08 |
| |
|
F123600 (P20) |
Min. 2 |
± 0.10 |
≤0. 03 |
± 0.20 |
≤0. 10 |
|
|
5 |
± 0.10 |
≤0. 04 |
± 0.20 |
≤0. 10 |
| |
|
10 |
± 0.10 |
≤0. 05 |
± 0.20 |
≤0. 10 |
| |
|
Max. 20 |
± 0.20 |
≤0. 06 |
± 0.20 |
≤0. 10 |
| |
|
F123615 (P100) |
Min. 20 |
± 0.35 |
≤0. 10 |
± 0.80 |
≤0. 30 |
|
|
50 |
± 0.40 |
≤0.12 |
± 0.80 |
≤0. 30 |
| |
|
Max. 100 |
± 0.80 |
≤0.15 |
± 0.80 |
≤0. 30 |
| |
|
F123601 (P200) |
Min. 50 |
± 0.50 |
≤0. 20 |
± 1.60 |
≤0. 60 |
|
|
100 |
± 0.80 |
≤0. 25 |
± 1.60 |
≤0. 60 |
| |
|
Max. 200 |
± 1.60 |
≤0. 30 |
± 1.60 |
≤0. 60 |
| |
|
F123602 (P1000) |
Min. 200 |
± 3 |
≤0.6 |
± 8 |
≤3 |
|
|
500 |
± 4 |
≤1.0 |
± 8 |
≤3 |
| |
|
Max. 1000 |
± 8 |
≤1.5 |
± 8 |
≤3 |
| |
|
F123603 (P5000) |
Min. 1000 |
± 12 |
≤3 |
± 40 |
≤15 |
|
|
2000 |
± 12 |
≤5 |
± 40 |
≤15 |
| |
|
Max. 5000 |
± 30 |
≤8 |
± 40 |
≤15 |
| |
|
F161201 (P10ml) |
Min. 1 mL |
± 30 |
≤6 |
± 60 |
≤30 |
|
|
2 mL |
± 30 |
≤6 |
± 60 |
≤30 |
| |
|
5 mL |
± 40 |
≤10 |
± 60 |
≤30 |
| |
|
Max. 10 mL |
± 60 |
≤16 |
± 60 |
≤30 |
| |
|
|
|
|
|
|
|
|
* P5000 과 P10ml 에는 Tip-ejector button 이 장착되어 있지 않으므로 이 점 유의 바랍니다 *